NOTE By default the Trigger is set to Internal and will continously trigger based on a time interval setting This may make it more difficult to make adjustments while setting up the sensor The best practice is as fol lows Go to the Main Menu Imager Trigger menu and select External Make sure there is no external trigger input Use the Trigger icon in the lowerright of the screen to manually trigger the sensor to capture an image as you set up and test Capture images of a range of samples to set up from the worst good part to the best bad part Start the setup with a good part Normally each part to be tested will be centered in the Field of View FOV 1 Teach the sensor a good reference part a Adjust the Region of Interest ROIThe ROI is a bluedotted box as shown below b Click Anywhere within the ROI to select it When selected the ROI is red with resize and rotational icons in the corners c Resize and move the ROI so that the feature of interest Resize the ROI so that it surrounds just the feature of interest In the Demo example the feature of interest is the Banner logo as shown below iVu Plus Users Manual 58 wwwbannerengineeringcom tel 7635443164 Online Only rev A